Sphygmomanometers for Clinical Use

 

Automated Devices for Clinical Use

Device Mode AAMI BHS ESH Circumstance Recommendation
A&D TM-2564G Osc Pass A/A   Serious protocol violations Questionable
BpTRU
BPM-100
BPM-300
Osc Pass A/A   At rest. Pressure ranges Recommended
Pass     Children Questionable
Pass A/A   Mixed adult/paediatric population
Protocol Violations
Should have used 1990 instead of 1993 BHS criteria but still A/A
Questionable
BpTRU
(unspecified)
        Pass Ad Hoc Protocol;
Abstract
Questionable
Colin 8800C Osc Pass     AAMI only Questionable
CVProfilor DO-2020 Dynamic linear deflation, Osc & Acoustic       New methodology, Passed Ad-hoc protocol Questionable
Datascope Accutorr Plus Osc Pass A/A   At rest Recommended
Pass     Trimline cuff replacement, at rest, protocol variation Questionable
    Pass Children. Adaptation of protocol along BHS guidelines. Recommended
DiaTecne
PulsePen
Tonometry       Abstract;
Pass ad hoc protocol;
Developer assessment
Questionable
Dinamap 1846 XT Osc Pass C/C   At rest; varying pressures, Recruitment violation, procedure violation Questionable
Dinamap ProCare Osc     Pass At rest Recommended
Osc Pass A/A   At rest Recommended
Dinamap ProCare Osc     Pass At rest Recommended
G-Care Electronics
SP-800
Osc     Pass Abstract Only Questionable
HoMedics BPA-300 Aus & Osc Pass     Sample size well over 85. Low mean BP Questionable
Lohmeier B-606 Osc     Pass Abstract Questionable
Omron 705IT
(See also SBPM)
Osc     Pass   Recommended
Pass A/A   Minor recruitment modification Recommended
    Pass Children and adolescents. Reasonable adaptation of protocol. Recommended
Omron 725 CIC Osc       Ad Hoc Protocol. Pointless study None
Omron 907 Osc     Pass Data analysed according to IP proposal only Questionable
Pass     At rest; AAMI only, Recruitment violation Questionable
Osc & Aus     Pass Elderly Population. Mode not stated. Recommended
Spengler Tensionic Mod EPS 112 Osc Pass B/A   At rest; Abstract Questionable
Tensoval duo control Aus & Osc Pass A/A Pass Auscultatory validation only Recommended
Terumo ES-H55 Osc
Double-cuff
Pass     Manufacturer Assessment Questionable
Welch Allyn Spot Vital Signs Osc Pass?     Only Method 1 completed. Ages 5-77 None
Welch Allyn 'Vital Signs' Osc Pass A/A   At rest. Pressure ranges Recommended
Fail C/D   At rest Not Recommended
    Fail Children. Adaptation of protocol along BHS guidelines. Not Recommended
 