Sphygmomanometers for Self-measurement of Blood Pressure (SBPM)
| UPPER ARM DEVICES | WRIST DEVICES | COMMUNITY DEVICES |
The following tables are lists of currently available Upper Arm and Wrist devices for Self-measurement of Blood Pressure and Devices for Measuring Blood Pressure in the Community. Discontinued devices are shown on a separate table. A complete list of all devices is available on our Device Index.
Upper Arm Devices for Self-measurement of Blood Pressure
| Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
Circumstance |
Recommendation Ref |
|
| A&D UA-631 (UA-779 Life Source) |
Osc |
|
|
Pass |
|
At rest, Recruitment violations |
Recommended 4 |
|
| A&D UA-704 |
Osc |
|
A/A |
|
|
Study details omitted |
Questionable 8 |
|
| A&D UA-705 |
Osc |
|
A/A |
|
|
At rest |
Recommended 18 |
|
| A&D UA-767 |
Osc |
Pass |
A/A |
|
|
At rest; not high BP |
Recommended 3 |
|
A&D UA-774 (UA-767 Plus) |
Osc |
|
A/A |
|
|
At rest; tables incomplete |
Recommended 11 |
|
| |
A/A |
|
|
At rest; Recruitment violations; Simultaneous readings |
Questionable 2 |
|
| A&D UA-778 |
Osc |
Pass |
B/A |
|
|
Children Only, Ad hoc protocol adaptation. (No general validation) |
Recommended 53 |
|
| A&D UA-787 |
Osc |
|
|
Pass |
|
|
Recommended 6 |
|
| A&D UA-853 |
Osc |
|
|
Pass |
|
Minor recruitment modification |
Recommended 46 |
|
| Andon KD-391 |
Osc (Semi-Auto) |
|
|
Pass |
|
At rest |
Recommended 56 |
|
| Andon KD-556 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E42 |
|
| Andon KD-575 |
Osc |
|
|
Fail |
|
At rest |
Not Recommended 59 |
|
| Andon KD-5031 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E43 |
|
| Andon KD-5913 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E45 |
|
| Andon KD-5915 |
Osc |
|
|
Pass |
|
At rest |
Recommended 57 |
|
| Andon KD-5963 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E44 |
|
| Andon MD-534 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E46 |
|
| Beurer BM35 |
Osc |
|
|
Pass |
|
HL-868BA Equivalence |
Recommended E67 |
|
| Beurer BM44 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E68 |
|
| Beurer BM58 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E78 |
|
| Beurer BM60 |
Osc |
|
|
Pass |
|
DS-400 Equivalence |
Recommended E69 |
|
| Beurer BM65 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E79 |
|
| Beurer SBM43 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E80 |
|
| Beurer SBM45 |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E81 |
|
| BIOS BD204 |
Osc |
|
|
Pass |
|
LD-578 Equivalence |
Recommended E74 |
|
| BIOS BD209 |
Osc |
|
|
Pass |
|
LD-578 Equivalence |
Recommended E66 |
|
| Braun BP4600 |
Osc |
|
|
Pass |
|
HL-868BA Equivalence |
Recommended E35 |
|
| Braun BP4900 |
Osc |
|
|
Pass |
|
HL-868BA Equivalence |
Recommended E40 |
|
| Braun BP5900 |
Osc |
|
|
Pass |
|
HL-868BA Equivalence |
Recommended E41 |
|
| Braun BP6000 |
Osc |
|
|
Pass |
|
TMB-986 Equivalence |
Recommended E95 |
|
| Braun BP6100 |
Osc |
|
|
Pass |
|
TMB-986 Equivalence |
Recommended E96 |
|
| Braun BP6200 |
Osc |
|
|
Pass |
|
TMB-986 Equivalence |
Recommended E97 |
|
| Citizen CH-432B |
Osc |
|
|
Pass |
|
At rest |
Recommended 40 |
|
| Citizen CH-452 |
Osc |
|
|
Pass |
|
CH-432B Equivalence |
Recommended E73 |
|
| Colson MAM BP3AA1-2 |
Osc |
|
|
Pass |
|
At rest |
Recommended 16 |
|
| Foracare Fora D 30 |
Osc |
|
|
Pass |
|
At rest |
Recommended 72 |
|
| Foracare Fora D 40b |
Osc |
|
|
|
Pass |
At rest |
Recommended 75 |
|
| Foracare Fora P30 Plus |
Osc |
|
|
|
Pass |
At rest |
Recommended 82 |
|
| Hartmann Tensoval duo control |
Aus-Osc |
Pass |
A/A |
Pass |
|
At rest |
Recommended 45 |
|
| Health & Life HL-868BA |
Osc |
|
|
Pass |
|
At rest |
Recommended 44 |
|
| Health & Life HL-888HA |
Osc |
|
Pass |
|
|
Abstract; Protocol adjustments |
Questionable 21 |
|
| |
A/A |
|
|
At rest |
Recommended 26 |
|
| Honsun LD-30 |
Osc |
|
|
Pass |
|
LD-578 Equivalence |
Recommended E66 |
|
| Honsun LD-578 |
Osc |
|
|
Pass |
|
At rest |
Recommended 47 |
|
IEM Stabil-O-Graph |
Osc |
|
A/A |
Pass |
|
At rest |
Recommended 27 |
|
| iHealth BP3 (KD-931) |
Osc |
|
|
|
Pass |
At rest |
Recommended 70 |
|
| |
|
Pass |
|
KD-5915 Equivalence |
Recommended E60 |
|
| Kingyield BP101H |
Osc |
|
|
Pass |
|
At rest |
Recommended 60 |
|
| Lloyds Pharmacy BP11 |
Osc |
|
|
Pass |
|
At rest |
Recommended 58 |
|
| Medel Check (92187) |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E64 |
|
| Medel Elite (92125) |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E65 |
|
| Medel Idea (91913 IDEA) |
Osc |
Pass |
A/A |
Pass |
|
SAA-102 Equivalence |
Recommended E33 |
|
| Medipro MediCare 100f |
Osc |
|
|
|
Pass |
At rest (Note: Not IBP Medipro 100f) |
Recommended 68 |
|
| Microlife BP 3AC1-1 |
Osc |
|
|
Pass |
|
At rest |
Recommended 15 |
|
| Pass |
A/A |
|
|
Pregnancy; Abstract |
Questionable 20 |
|
| Pass |
B/A |
|
|
Pre-eclampsia; Abstract |
Questionable 20 |
|
| Microlife BP 3AC1-1 PC |
Osc |
|
|
Pass |
|
BP 3AC1-1 Equivalence |
Recommended E1 |
|
| |
|
Pass |
|
ESRD population |
Recommended 28 |
|
| Microlife BP 3AC1-2 |
Osc |
|
|
Pass |
|
BP 3AC1-1 Equivalence |
Recommended E2 |
|
| Microlife BP 3AG1 |
Osc |
|
A/A |
|
|
BP 3BT0-A Equivalence |
Recommended E3 |
|
| Microlife BP 3BTO-1 |
Osc |
|
A/A |
|
|
BP 3BT0-A Equivalence |
Recommended E4 |
|
| Microlife BP 3BTO-A |
Osc |
|
A/A |
|
|
Small recruitment violation |
Recommended 7 |
|
| Pass |
A/B |
|
|
Normotensive Pregnancy |
Recommended 12 |
|
| Pass |
B/B |
|
|
Non-proteinuric HBP |
| Pass |
A/B |
|
|
Pre-Eclampsia |
| Microlife BP 3BTO-A(2) |
Osc |
|
A/A |
|
|
BP 3BT0-A Equivalence |
Recommended E5 |
|
| Microlife BP 3BTO-AP |
Osc |
|
A/A |
|
|
BP 3BT0-A Equivalence |
Recommended E6 |
|
| Microlife BP A100 |
Osc |
|
|
Pass |
|
BP A100 Plus Equivalence |
Recommended E7 |
|
| |
|
Pass |
|
With "M-L Size" Cuff |
Recommended FTS, 52 |
|
| Microlife BP A100 Plus |
Osc |
|
|
Pass |
|
At rest |
Recommended 17 |
|
| |
|
Pass |
|
At rest |
Recommended 29 |
|
| Microlife Exact BP (BP3MD1-3) |
Osc |
|
|
Pass |
|
BP A100 Equivalence |
Recommended E21 |
|
| Microlife RM 100 |
Osc |
|
A/A |
|
|
BP 3BT0-A Equivalence |
Recommended E8 |
|
| Microlife WatchBP Home (BP3MX1-1) |
Osc |
|
|
Pass |
|
At rest |
Recommended 25 |
|
| |
A/A |
|
|
Pregnancy |
Recommended 66 |
|
| |
B/A |
|
|
Pre-Eclampsia |
Recommended 66 |
|
| Microlife WatchBP Home A (BP3MX1-3) |
Osc |
|
|
Pass |
|
Watch BP Home Equivalence |
Recommended E27 |
|
| Microlife WatchBP Home N (BP3MX1-4) |
Osc |
|
|
Pass |
|
Watch BP Home Equivalence |
Recommended E70 |
|
| Microlife WatchBP Home S (BP3MX1-5) |
Osc |
|
|
Pass |
|
Watch BP Home Equivalence |
Recommended E28 |
|
| Microlife WatchBP O3 (3MZ0) |
Osc |
|
|
Pass |
|
At rest |
Recommended 61 |
|
| Microlife WatchBP O3 (3MZ1) |
Osc |
|
|
Pass |
|
At rest |
Recommended 62 |
|
| Nissei DS-400 |
Osc |
|
|
Pass |
|
At rest |
Recommended 48 |
|
| Nissei DS-500 |
Osc |
|
|
Pass |
|
DS-400 Equivalence |
Recommended E38 |
|
| Nissei DSK-1011 |
Osc |
|
|
|
Pass |
At rest |
Recommended 76 |
|
| Nissei DSK-1031 |
Osc |
|
|
|
Pass |
At rest |
Recommended 77 |
|
| Nissei DSK-1051 |
Osc |
|
|
|
Pass |
DSK-1031 Equivalence |
Recommended E98 |
|
Omron 705IT (HEM-759-E) |
Osc |
|
|
Pass |
|
|
Recommended 9 |
|
| Pass |
A/A |
|
|
Minor recruitment modification |
Recommended 14 |
|
| |
|
Pass |
|
Children and adolescents. Reasonable adaptation of protocol. |
Recommended 22 |
|
Omron 708-BT (HEM-7081-ITE) |
Osc |
Pass |
A/A |
|
|
M7 (HEM-780-E) Equivalence |
Recommended E103 |
 |
Omron BP760 (HEM-7220-Z) |
Osc |
|
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence New sensor |
Recommended E87 |
|
Omron BP785 (HEM-7222-Z) |
Osc |
|
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence New sensor |
Recommended E88 |
|
Omron HEM-705CP-II (HEM-759-E2) |
Osc |
|
|
Pass |
|
705IT (HEM-759-E) Equivalence |
Recommended E12 |
|
Omron HEM-746C (HEM-746C-SH) |
Osc |
|
|
Pass |
|
M2 Compact (HEM-7102-E) Equivalence |
Recommended E54 |
|
| Omron HEM-780REL |
Osc |
|
|
Pass |
|
At rest; "East Wrap" cuff. Poor study and paper |
Questionable 32 |
|
Omron HEM-1020 |
Osc |
|
|
Pass |
|
Supplementary Results |
Recommended 71 |
|
Omron HEM-7051 (HEM-7051-SH) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E48 |
|
| Omron HEM-7080IC |
Osc |
Pass |
A/A |
Pass |
|
705IT (HEM-759-E) Equivalences |
Recommended E20 |
|
Omron HEM-7101 (HEM-7101-SH) |
Osc |
|
|
Pass |
|
M2 Compact (HEM-7102-E) Equivalence |
Recommended E53 |
|
| Omron HEM-7250-IT |
Osc |
|
|
|
Pass |
HEM-7251G Equivalence |
Recommended E102 |
 |
| Omron HEM-7251G |
Osc |
|
|
|
Pass |
At rest |
Recommended 83 |
|
Omron HEM-Solar (HEM-4500-SOLE) |
Osc |
|
|
Pass |
|
M1 Plus (HEM-4011C-E) Equivalence |
Recommended E26 |
|
Omron IA2 (HEM-7011-C1) |
Osc |
Pass |
A/A |
Pass |
|
705IT (HEM-759-E) Equivalence |
Recommended E13 |
|
Omron i-C10 (HEM-7070-E) |
Osc |
|
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence |
Recommended E18 |
|
| Pass |
A/A |
|
|
M7 (HEM-780-E) Equivalence |
Recommended E19 |
|
Omron i-Q132 SpotArm (HEM-1010-E) |
Osc |
|
|
Pass |
|
Supplementary Results |
Recommended 71 |
|
Omron i-Q142 SpotArm (HEM-1040-E) |
Osc |
|
|
|
Pass |
At rest |
Recommended 81 |
|
Omron M1 (HEM-4030-E) |
Osc |
|
|
Pass |
|
M1 Plus (HEM-4011C-E) Equivalence |
Recommended E59 |
|
Omron M1 Classic (HEM-442-E) |
Osc |
|
|
Pass |
|
M1 Plus (HEM-4011C-E) Equivalence |
Recommended E10 |
|
Omron M1 Compact (HEM-4022-E) |
Osc |
|
|
Pass |
|
M1 Plus (HEM-4011C-E) Equivalence |
Recommended E11 |
|
Omron M1 Plus (HEM-4011C-E) |
Osc |
|
|
Pass |
|
At rest |
Recommended 29 |
|
Omron M2 (HEM-7117-E) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E49 |
|
Omron M2 (HEM-7117-E8) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E83 |
|
Omron M2 (HEM-7119-E) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E90 |
|
Omron M2 Basic (HEM-7116-E) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E50 |
|
Omron M2 Basic (HEM-7116-E2) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E89 |
|
Omron M2 Basic (HEM-7116-E8) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E82 |
|
Omron M2 Compact (HEM-7102-E) |
Osc |
|
|
Pass |
|
At rest |
Recommended 31 |
|
Omron M3 (HEM-7200-E) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E51 |
|
| |
|
|
Pass |
Supplementary Results |
Recommended 65 |
|
Omron M3 (HEM-7200-E2) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E91 |
|
Omron M3 (HEM-7200-E8) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence New sensor |
Recommended E84 |
|
Omron M3 Intellisense (HEM-7051-E) |
Osc |
|
|
Pass |
|
At rest |
Recommended 31 |
|
Omron M3W (HEM-7202-E) |
Osc |
|
|
|
Pass |
At rest |
Recommended 78 |
|
Omron M4-I (HEM-752-E) |
Osc |
Pass |
A/A |
Pass |
|
705IT (HEM-759-E) Equivalence |
Recommended E15 |
|
Omron M5-I (HEM-757-E) |
Osc |
|
|
Pass |
|
|
Recommended 9 |
|
| |
|
Pass |
|
Elderly Population |
Recommended 33 |
|
Omron M6 (HEM-7001-E) |
Osc |
|
|
Pass |
|
At rest |
Recommended 19 |
|
| |
|
Pass |
|
General Population |
Recommended 34 |
|
| |
|
Pass |
|
Obese Population |
Recommended 34 |
|
| |
|
Pass |
|
Elderly Population |
Recommended 23 |
|
| Pass |
A/A |
|
|
705IT (HEM-759-E) Equivalence |
Recommended E23 |
|
Omron M6 (HEM-7211-E) |
Osc |
Pass |
A/A |
Pass |
|
705IT (HEM-759-E) Equivalences |
Recommended E58 |
|
| |
|
|
Pass |
Supplementary Results |
Recommended 65 |
|
Omron M6 (HEM-7211-E8) |
Osc |
|
|
Pass |
|
705IT (HEM-759-E) Equivalence New sensor |
Recommended E85 |
|
Omron M6 Comfort (HEM-7000-E) |
Osc |
|
|
Pass |
|
At rest |
Recommended 29 |
|
| Pass |
A/A |
|
|
M7 (HEM-780-E) Equivalence |
Recommended E17 |
|
Omron M6 Comfort (HEM-7221-E) |
Osc |
|
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence |
Recommended E56 |
|
| Pass |
A/A |
|
|
M7 (HEM-780-E) Equivalence |
Recommended E57 |
|
Omron M6 Comfort (HEM-7221-E8) |
Osc |
|
|
|
Pass |
At rest |
Recommended 80 |
|
| |
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence New sensor |
Recommended E86 |
|
Omron M6 Comfort (HEM-7223-E) |
Osc |
|
|
|
Pass |
M6 Comfort (HEM-7221-E8) Equivalence |
Recommended E92 |
|
Omron M6W (HEM-7213-E) |
Osc |
|
|
|
Pass |
At rest |
Recommended 79 |
|
Omron M7 (HEM-780-E) |
Osc |
Pass |
A/A |
|
|
General population |
Recommended 38 |
|
| |
|
Pass |
|
Population requiring Large Cuff |
Recommended 35 |
|
| |
A/A |
|
|
Pregnancy |
Recommended 50 |
|
| |
B/B |
|
|
Pre-Eclampsia |
Recommended 50 |
|
Omron M10-IT (HEM-7080IT-E) |
Osc |
|
|
Pass |
|
M6 Comfort (HEM-7000-E) Equivalence |
Recommended E16 |
|
| Pass |
A/A |
|
|
M7 (HEM-780-E) Equivalence |
Recommended E9 |
|
Omron MX2 Basic (HEM-742-E2) |
Osc |
|
|
Pass |
|
M2 Compact (HEM-7102-E) Equivalence |
Recommended E55 |
|
Omron MX3 Plus (HEM-742-E) |
Osc |
|
|
Pass |
|
Subject to corrected tables being published |
Recommended 13, FTS |
|
Omron SEM-1 (HEM-7051-C12) |
Osc |
|
|
Pass |
|
M3 Intellisense (HEM-7051-E) Equivalence |
Recommended E52 |
|
| Oregon Scientific BPU 330 |
Osc |
|
|
Pass |
|
At rest, Plots incorrect |
Recommended 49 |
|
| Oregon Scientific BPW810 |
|
|
|
Pass |
|
At rest |
Questionable 24 |
|
| Panasonic EW3106 |
Osc |
|
|
Pass |
|
At rest |
Recommended 69, FTS |
|
| Panasonic EW3109 |
Osc |
|
|
Pass |
|
At rest |
Recommended 69, FTS |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Classic Check (100) |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E30 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Classic Check (200) |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E72 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Comfort Check |
Osc |
|
|
Pass |
|
At rest |
Recommended 54 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) CS 410 |
Osc |
|
|
Pass |
|
At rest |
Recommended 41 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) CSI 610 |
Osc |
|
|
Pass |
|
At rest |
Recommended 39 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Daily Check |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E75 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Digit Extra |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E93 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Digit Smart |
Osc |
|
|
Pass |
|
KD-5915 Equivalence |
Recommended E94 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Help Check |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E76 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Mamy Check |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E77 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) My Check |
Osc |
|
|
Pass |
|
At rest |
Recommended 54 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) One Check |
Osc |
|
|
Pass |
|
My Check Equivalence |
Recommended E31 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Personal Check |
Osc |
|
|
Pass |
|
At rest |
Recommended 54 |
|
Polygreen K-Jump KP-7770 |
Osc |
|
|
Pass |
|
K-Jump KP7500D Equivalence |
Recommended E71 |
|
| Rossmax ME 701 series |
Osc |
|
A/A |
Pass |
|
At rest |
Recommended 30 |
|
| Seinex SE-9400 |
Osc |
|
|
Pass |
|
At rest |
Recommended 15 |
|
| Sensacare SAA-102 (Standard) |
Osc |
Pass |
A/A |
Pass |
|
Three studies. BHS under reported. |
Recommended 36 |
|
Spengler K-Jump KP7500D |
Osc |
|
|
Pass |
|
At rest |
Recommended 29 |
|
Spengler K-Jump Pro M |
Osc |
|
|
Pass |
|
At rest |
Recommended 42 |
|
| Transtek TMB-986 |
Osc |
|
|
Pass |
|
At rest |
Recommended 67 |
|
| UEBE Visomat Comfort 20/40 |
Osc |
|
|
Pass |
|
At Rest |
Recommended 43 |
|
| UEBE Visomat Comfort Eco |
Osc |
|
|
Pass |
|
At rest |
Recommended 51 |
|
| UEBE Visomat Comfort Form |
Osc |
|
|
Pass |
|
At rest |
Recommended 55 |
|
| UEBE Visomat Double Comfort |
Osc |
|
|
|
Pass |
At rest |
Recommended 64, 74 |
|
| Aus |
|
|
|
Pass |
At rest |
Recommended 63, 73 |
| Welch-Allyn transtelephonic home monitor |
Osc |
Pass |
? |
|
|
Parkinson's disease, 30 sub * 2 measurements |
Questionable 5 |
|
* Note: Grading criteria had not been established though BHS protocol was in operation.5
References for Validation of Upper Arm Devices for Self-measurement of Blood Pressure
| 1 |
Dieterle T, Battegay E, Bucheli B, Martina B. Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and wrist. Blood Press Moni 1998;3:339-346 |
| 2 |
Kobalava ZD, Kotovskaia IV, Rusakova OS, Babaeva LA. Validation of UA-767 Plus device for self measurement of blood pressure Clin. Pharmacol. Ther. 2003;12 (2):70-72 |
| 3 |
Rogoza AN, Pavlova TS, Sergeeva MV. Validation of A & D UA-767 device for the self-measurement of blood pressure. Blood Press Monit 2000;5:227-23d1 |
| 4 |
Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the A & D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002;7:243-248 |
| 5 |
Herbst T, Anis-Anwar Y, Giacco S, White W. Evaluation of the overall system precision of the Welch-Allyn transtelephonic home blood pressure monitor in adults with Parkinson's disease Blood Press Monit 2002;7:285-288 |
| 6 |
Longo D, Toffanin G, Garbelotto R, Zaetta V, Businaro L, Palatini P. Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2003;8:91-95 |
| 7 |
Cuckson AC, Reinders A, Shabeeh H, Shennan AH. Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol Blood Press Monit 2002;7(6):319-324 |
| 8 |
Koblava ZD, Kotovskaya YV, Rodionov E Validation of semi-automatic device UA-704 for self-measurement of blood pressure Blood Press Monit 2005;10:223-225 |
| 9 |
El Assad M, Topouchian J, Asmar R. Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT Blood Press Monit 2003;8(3):127-133 |
| 11 |
Verdecchia P, Angeli F, Poeta F, Reboldi GP, Borgioni C, Pittavini L, Porcellati C. Validation of the A&D UA774 (UA-767Plus) device for self measurement of blood pressure. Blood Press Monit 2004;9:225-229 |
| 12 |
Reinders A, Cuckson AC, Lee JTM, Shennan AH. An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A. BJOG 2005;112(7):915-920 |
| 13 |
Coleman A, Freeman P, Steel S, Shennan A Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol Blood Press Monit 2005;10:165-168 |
| 14 |
Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Blood Press Monit 2006;11:27-32 |
| 15 |
Topouchian JA, El Assad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two devices for self-measurement of brachial blood pressure according to the International Protocol of the European Society of Hypertension: the Seinex SE-9400 and the Microlife BP 3AC1-1 Blood Press Monit 2005;10:325-331 |
| 16 |
Pereira T, Maldonado J. Validação de aparelho de medição automática da pressão arterial Colson MAM BP3AA1-2 segundo o Protocolo da Sociedade Europeia de Hipertensão Rev Port Cardiol 2005;24(11):1341-1351 |
| 17 |
Stergiou GS, Giovas PP, Neofytou MS, Adamopoulos DN. Validation of the Microlife BP A 100 Plus device for self-home blood pressure measurement according to the International Protocol Blood Press Monit 2006;11:157-160 |
| 18 |
Verdecchia P, Angeli F, Gattobigio R, Angeli E, Pede S, Pittavini L. Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Blood Press Monit 2006;11:223-227 |
| 19 |
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) Blood Press Monit 2006;11:165-171 |
| 20 |
de Greeff A, Reggiori F, Anthony J, Shennan A. The Microlife 3AC1: An accurate blood pressure measurement device in pregnancy and pre-eclampsia J Hypertens 2006;24 (suppl 4):S279 Abstract P14.475 |
| 21 |
Tholl U, Assouline E. Validation of the automatic arm type non-invasive blood pressure monitor HL-888HA J Hypertens 2006;24 (suppl 4):S285 Abstract P14.496 |
| 22 |
Stergiou GS, Yiannes NG, Rarra VC. Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children and adolescents: The Arsakion School Study Blood Press Monit 2006,11:229-234 |
| 23 |
Altunkan S, Iliman N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper arm blood pressure measuring device according to the International Protocol in elderly patients Blood Press Monit 2008;13:117-122 |
| 24 |
Li L, Wang Hui W, Jin Lan W. Validation of the Oregon Scientific BPW810 blood pressure monitor according to the European Society of Hypertension Protocol. Blood Press Monit 2006;11:343-347 |
| 25 |
Stergiou GS, Giovas PP, Charilaos P. Gkinos CP, Patouras JD. Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol Blood Press Monit 2007;12:185-188 |
| 26 |
de Greeff A, Shennan AH. The Health and Life device (HL888HA): accuracy assessment in an adult population according to the British Hypertension Society protocol Blood Press Monit 2007;12:107-111 |
| 27 |
Westhoff TH, Schmidt S, Zidek W, van der Giet M. Validation of the Stabil-O-Graph blood pressure self-measurement device Journal of Human Hypertension 2007;22,233-235,Research Letter |
| 28 |
Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG. Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy Blood Press Monit 2007;12:227-232 |
| 29 |
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension Vascular Health and Risk Management 2007;3(4):389-400 |
| 30 |
de Greeff A, Shennan AH. The Rossmax (ME 701 series) upper arm device: accuracy assessment in an adult population according to the International and the British Hypertension Society protocols Blood Press Monit 2008;13:43-48 |
| 31 |
Asmar R, Khabouth J, Topouchian J, El Feghali R, Mattar J. Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Press Monit 2010;15(1):49-54 |
| 32 |
Viera AJ, Hinderliter AL. Validation of the HEM-780REL with easy wrap cuff for self-measurement of blood pressure according to the European Society of Hypertension International Protocol Blood Press Monit 2007;12:335-338 |
| 33 |
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension Blood Press Monit 2007;12:233-242 |
| 34 |
Altunkan S, Iliman N, Kayaturk N, Altunkan E. Validation of the Omron M6 (HEM-7001-E) upper-arm blood pressure measuring device according to the International Protocol in adults and obese adults Blood Press Monit 2007;12:219-225 |
| 35 |
El Feghali RN, Topouchian JA, Pannier BM, El Assaad HA, Asmar RG. Validation of the Omron M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension Blood Press Monit 2007;12:173-178 |
| 36 |
Zaetta V, Daniele L, Perkovic D, Prattico F, Barisa M, Perfetti P, Gabrieli A, Buonocore F, Winnicki M. Validation of the SAA-102 home blood pressure monitor according to the protocols of the European Society of Hypertension, the Association for the Advancement of Medical Instrumentation and the British Society of Hypertension Blood Press Monit 2007;12:363-368 |
| 38 |
Coleman A, Steel S, Freeman P, de Greeff A, Shennan A. Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit 2008;13:49-54 |
| 39 |
Pini C, Pastori M, Baccheschi J, Omboni S, Parati G. Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension Blood Press Monit 2007;12:179–184 |
| 40 |
Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C Blood Press Monit 2008;13:55-62 |
| 41 |
Pini C, Natalizi A, Gerosa PF, Frigerio M, Omboni S, Parati G. Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension Blood Press Monit 2008;13:177-182 |
| 42 |
de Greeff A, Ahmed N, Sheppard E, Andrew H. Shennan AH. A self-measurement device suitable for hospital use: validation of the Spengler Pro M according to the International Protocol of the European Society of Hypertension Blood Press Monit 2008;13:183-186 |
| 43 |
Stergiou GS, Tzamouranis D, Nasothimiou EG, Protogerou AD. Can an electronic device with a single cuff be accurate in a wide range of arm size? Validation of the Visomat Comfort 20/40 device for home blood pressure monitoring Journal of Human Hypertension 2008;23:Advanced Online Publication |
| 44 |
Chen HE, Cui Y, Sheng CS, Li LH, Li Y, Wang JG. Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2008;13:305-308 |
| 45 |
de Greeff A, Arora J, Hervey S, Liu B, Shennan AH. Accuracy assessment of the Tensoval duo control according to the British and European Hypertension Societies' standards Blood Press Monit 2008;13:111-116 |
| 46 |
Bonso E, Ragazzo F, Palatini P. Validation of A&D UA-85X device for blood pressure measurement. Blood Press Monit 2008;13:339-341 |
| 47 |
Zhang Y, Wang J, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2009;14:128-131 |
| 48 |
Duhig KE, de Greeff A, van der Westhuizen A, Baker E, Shennan AH. Validation of the Nissei DS-400 in a low-resource setting. Blood Press Monit 2009;14:132-135 |
| 49 |
Li L, Zhang XY, Yan CH, Liang QX. Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol. Vascular Health and Risk Management 2008;4(5):1121-1125 |
| 50 |
de Greeff A, Beg Z, Gangji Z, Dorney E, Shennan AH. Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7. Blood Press Monit 2009;14:37-40 |
| 51 |
Perkovic D, Guarnieri C, Palatini P. Validation of the visomat comfort eco blood pressure measuring device according to the International Protocol. Blood Press Monit 2009;14:178-180 |
| 52 |
Bonso E, Dorigatti F, Palatini P. Accuracy of the BP A100 blood pressure measuring device coupled with a single cuff with standard-size bladder over a wide range of arm circumferences. Blood Press Monit 2009;14:216-219 |
| 53 |
Narogan MV, Narogan MI, Syutkina EV. Validation of A&D UA-778 blood pressure monitor in children. Blood Press Monit 2009;14:228-231 |
| 54 |
Germano G, Psimenos A, Sarullo F, Venditti A, Pecchioli V, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol: The Pic Indolor Personal Check, Comfort Check, My Check and Travel Check. Blood Pressure 2009;18 (Suppl 1):15-23 |
| 55 |
Bonso E, Dorigatti F, Saladini F, Palatini P. Validation of the Visomat Comfort form device for home blood pressure measurement according to the International Protocol. Blood Press Monit 2010;15(1):63-66 |
| 56 |
Wan Y, Pan F, Liu Y, Liang Y, Yang Z, Zhao H, Xu Y. Validation of the Andon KD-391 semiautomated blood pressure monitor in adults according to the International Protocol. Blood Press Monit 2009;14:181-184 |
| 57 |
Huang QF, Wang J, Sheng CS, Zhang NN, Li Y, Wang JG. Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol. Blood Press Monit 2010;15(4):232-234 |
| 58 |
Coleman A, Steel S, deGreeff A, Shennan A. Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension. Blood Press Monit 2010;15(3):163-6. |
| 59 |
Pan F, Wan Y, Chen W, Li N, Tong W, Xu Y. Validation of the Andon KD-575 automated blood pressure monitor according to the European Society of Hypertension International protocol in adults. Blood Press Monit 2010;15(3):167-70. |
| 60 |
Zeng WF, Huang QF, Sheng GS, Li Y, Wang JG. Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol Blood Press Monit 2010;15(6):329-31. |
| 61 |
Stergiou GS, Tzamouranis D, Nasothimiou EG, Karpettas N, Protogerou A. Are there really differences between home and daytime ambulatory blood pressure? Comparison using a novel dual-mode ambulatory and home monitor. J Hum Hypertens 2010;24(3):207-12. |
| 62 |
Ragazzo F, Saladini F, Palatini P. Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol. Blood Press Monit 2010;15(1):59-62. |
| 63 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2011;16(4):208-210. |
| 64 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometric mode, for clinic use and self-measurement in a general population according to the European Society of Hypertension International Protocol, revision 2010. Blood Press Monit 2011;16(5):262-264 |
| 65 |
Topouchian J, Agnoletti D, Blacher J, Youssef A, Ibanez I, Khabouth J, Khawaja S, Beaino L, Asmar R. Validation of four automatic devices for selfmeasurement of blood pressure according to the international protocol of the European Society of Hypertension. Vascular Health and Risk Management 2011:7;709-717 doi: 10.2147/VHRM.S27193 |
| 66 |
Chung Y, de Greeff A, Shennan A. Validation and Compliance of a Home Monitoring Device in Pregnancy: Microlife WatchBP Home. Hypertens Pregnancy 2009;28(3):348-59. |
| 67 |
Liu WJ, Li SG, Song Z, Gong W. Validation of the TRANSTEK blood pressure monitor TMB-986 for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2010;15(5):278-80 doi: 10.1097/MBP.0b013e32833e43ca |
| 68 |
Yi J, Wan Y, Pan F, Yu X, Zhao H, Shang F, Xu Y. Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2011;16(4):197-199. doi: 10.1097/MBP.0b013e328346d715 |
| 69 |
Bonso E, Dorigatti F, Palatini P. Validation of Panasonic EW3106 and EW3109 devices for blood pressure measurement according to the International Protocol. Blood Press Monit 2010;15(1):55-58. doi: 10.1097/MBP.0b013e32833531b3 |
| 70 |
Chen C, Shang F, Wang J, Chen J, Ji N, Wan Y. Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2012;17(6):253-256. doi: 10.1097/MBP.0b013e32835b9ecb |
| 71 |
Takahashi H, Yoshika M, Yokoi T. Validation of home blood pressure-monitoring devices, Omron HEM-1020 and Omron i-Q132 (HEM-1010-E), according to the European Society of Hypertension International Protocol. Blood Press Monit 2011;16(4):203-7. doi: 10.1097/MBP.0b013e328348b688 |
| 72 |
Palatini P. Validation of the Foracare Fora D30 upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol 2002 [Internet]. Dublin: dablEducational Trust; 2010 Apr 05. 6 p. Available from: http://www.dableducational.org/Publications/2010/ESH-IP 2002 Validation of Foracare Fora D30.pdf |
| 73 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Apr 27. 4 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of UEBE Visomat_Double_Comfort (Auscultatory Mode).pdf. |
| 74 |
Masiero S, Fania C, Palatini P. Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Apr 27. 4 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of UEBE Visomat_Double_Comfort (Oscillometric Mode).pdf. |
| 75 |
Ding D. Validation of the Foracare Fora D40b upper arm blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol Revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Jul 29. 7 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Foracare Fora D40b.pdf |
| 76 |
de Greeff A, Shennan AH. Validation of the Nissei DSK-1011 upper arm blood pressure monitor, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Sep 01. 9 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei DSK-1011.pdf. |
| 77 |
de Greeff A, Shennan AH. Validation of the Nissei DSK-1031 upper arm oscillometric blood pressure monitor intended for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Oct 28. 3 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei DSK-1031.pdf. |
| 78 |
Takahashi H. Validation of the Omron M3W upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Mar 01. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M3W (HEM-7202-E).pdf |
| 79 |
Takahashi H. Validation of the Omron M6W upper arm blood pressure monitor, in oscillometry mode, for self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Mar 01. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M6W (HEM-7213-E).pdf |
| 80 |
Asmar R. Validation of the OMRON M6 COMFORT (HEM-7221-E8) Upper Arm blood pressure monitor, for self/home measurement, according to the revised (2010) European Society of Hypertension International Protocol. [Internet]. Dublin: dablEducational Trust; 2012 Jun 06. 3 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron M6 Comfort (HEM-7221-E8).pdf |
| 81 |
Takahashi H. Validation of the Omron i-Q142(HEM-1040-E), an upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Jun 07. 4 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Omron i-Q142 (HEM-1040-E).pdf |
| 82 |
Ding D. Validation of the FORA P30 Plus Upper Arm Blood Pressure Monitor, for Clinic Use and Self/Home Measurement, According to the European Society of Hypertension International Protocol Revision 2010 [Internet]. Dublin: dablEducational Trust; 2012 Oct 19. 6 p. Available from: http://www.dableducational.org/Publications/2012/ESH-IP 2010 Validation of Foracare Fora P30 Plus.pdf |
| 83 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON HEM-7251G upper arm blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron HEM-7251G.pdf |
Wrist Devices for Self-measurement of Blood Pressure
Recommendations for all wrist devices are based on their being positioned strictly according to manufacturer instructions.
For a complete list of devices see our Device Index and for discontinued devices see our Discontinued Devices Index.
New to the website |
|
|
| Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
Circumstance |
Recommendation Ref |
|
| A&D UB-511 |
Wrist |
|
Pass |
|
|
At rest |
Questionable 14 |
|
| Braun BP 3000 |
Wrist |
Pass |
|
|
|
At Rest |
Questionable 6 |
|
| Pass |
|
|
|
At rest |
Questionable 15 |
|
| Braun BP 3550 |
Wrist |
|
|
Pass |
|
At rest. Questionable in high range. |
Recommended 16 |
|
| Braun BP VitalScan Plus 1650 |
Wrist |
|
|
Pass |
|
At rest |
Recommended 8 |
|
| Braun BP VitalScan Plus 1700 |
Wrist |
|
|
Pass |
|
BP1650 Equivalence |
Recommended E24 |
|
| Braun BP VitalScan Plus 1750 |
Wrist |
|
|
Pass |
|
BP1650 Equivalence |
Recommended E25 |
|
| Braun PrecisionSensor BP2000 |
Wrist |
Pass |
|
|
|
At rest, Manufacturer Assistance |
Questionable 1 |
|
| Braun PrecisionSensor BP2550 |
Wrist |
Pass |
B/B |
|
|
Recruitment violation (AAMI) Recruitment violation, Analysis errors (BHS) |
Questionable 2, 3 |
|
| |
?B/B |
|
|
Abstract, SBP results unclear |
Questionable 5 |
|
| Braun PrecisionSensor BP2550 (UG) |
Wrist |
|
|
Pass |
|
At rest |
Recommended 7 |
|
| Braun VitalScan |
Wrist |
Pass |
?? |
|
|
Abstract; Version not stated |
Questionable 4 |
|
| Citizen CH-656C |
Wrist |
|
|
Pass |
|
At rest |
Recommended 9 |
|
| Health & Life HL168JD |
Wrist |
|
|
Pass |
|
At rest |
Recommended 24 |
|
| iHealth BP7 (KD-972) |
Wrist |
|
|
|
Pass |
At rest |
Recommended 36 |
 |
| Kingyield BP210 |
Wrist |
|
|
|
Pass |
Study (29) extended to ESH-IP 2010 requirements |
Recommended 32 |
|
| Medel Replay (91914REPLAY) |
Wrist |
Pass |
B/B |
Pass |
|
SAW-102 Equivalence |
Recommended E34 |
|
| Microlife BP W100 |
Wrist |
Pass |
A/A |
Pass |
|
Supine position |
Recommended 23 |
|
| Microlife BP W200-1 |
Wrist |
|
|
Pass |
|
MAM averaging and supine position |
Recommended 21 |
|
| Nissei WSK-1011 |
Wrist |
|
|
|
Pass |
At rest |
Recommended 31 |
|
| Nissei WSK-1021 |
Wrist |
|
|
|
Pass |
WSK-1011 Equivalence |
Recommended E99 |
|
Omron 637IT (HEM-637-E) |
Wrist |
Pass |
B/A |
Fail |
|
Abstract; BHS 'pass' incorrectly reported |
Questionable 12 |
|
| |
|
|
|
AD Hoc protocol, + position sensor, std. protocol projections inconclusive |
Questionable 11 |
| |
|
Pass |
|
Adults |
Recommended 12 |
|
| |
|
Pass |
|
Obese Adults |
Recommended 12 |
|
| |
|
Pass |
|
Elderly |
Recommended 10 |
|
Omron R3 Intellisense (HEM-6021-E) |
Wrist |
|
|
Pass |
|
R3-I Plus (HEM-6022-E) Equivalence |
Recommended E47 |
|
Omron R3-I Plus (HEM-6022-E) |
Wrist |
|
|
Pass |
|
At rest |
Recommended 26 |
|
Omron R5-I (HEM-630-E) |
Wrist |
|
|
Pass |
|
Elderly Population |
Recommended 18 |
|
Omron R6 (HEM-6000-E) |
Wrist |
|
|
Pass |
|
R7 (HEM-637-IT) Equivalence |
Recommended E22 |
|
Omron R6 (HEM-6000-E7) |
Wrist |
|
|
Pass |
|
R7 (HEM-637-IT) Equivalence New sensor |
Recommended E39 |
|
Omron R6 (HEM-6052-E) |
Wrist |
|
|
Pass |
|
R7 (HEM-637-IT) Equivalence |
Recommended E62 |
|
Omron R6 (HEM-6052-E7) |
Wrist |
|
|
Pass |
|
R7 (HEM-637-IT) Equivalence New sensor |
Recommended E63 |
|
Omron R7 (HEM-637-IT) |
Wrist |
|
|
Pass |
|
At rest |
Recommended 13 |
|
Omron R7 (HEM-637-E7) |
Wrist |
|
|
Pass |
|
R7 (HEM-637-IT) Equivalence New sensor |
Recommended E61 |
|
Omron RS1 (HEM-6120-E) |
Wrist |
|
|
|
Pass |
RS3 (HEM-6130-E) Equivalence |
Recommended E100 |
 |
Omron RS2 (HEM-6121-E) |
Wrist |
|
|
|
Pass |
RS3 (HEM-6130-E) Equivalence |
Recommended E101 |
 |
Omron RS3 (HEM-6130-E) |
Wrist |
|
|
|
Pass |
At rest |
Recommended 34 |
|
Omron RS6 (HEM-6221-E) |
Wrist |
|
|
|
Pass |
At rest |
Recommended 35 |
|
Omron RX3 (HEM-640-E) |
Wrist |
|
|
Pass |
|
Abstract only |
Questionable 20 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Self Check |
Wrist |
|
|
Pass |
|
Travel Check Equivalence |
Recommended E32 |
|
Pic Solution (Artsana) Pic Indolor (Artsana) Travel Check |
Wrist |
|
|
Pass |
|
At rest |
Recommended 25 |
|
| Rossmax S150 |
Wrist |
|
|
|
Pass |
Study (30) protocol violations corrected |
Recommended 33 |
|
| Samsung SBM-100A |
Wrist |
|
|
Fail |
|
At rest |
Not Recommended 17 |
|
| Sensacare SAW-102 (Standard) |
Wrist |
Pass |
B/B |
Pass |
|
At rest |
Recommended 19 |
|
| Transtek TMB-988 |
Wrist |
|
|
Pass |
|
At rest |
Recommended 28 |
|
| UEBE Visomat Handy |
Wrist |
|
|
Pass |
|
At rest |
Recommended 27 |
|
| Visicor HM40 |
Wrist |
|
|
Pass |
|
Supine position |
Recommended 22 |
|
References for Validation of Wrist Devices for Self-measurement of Blood Pressure
| 1 |
Wessig K, Hollinger S, Schmalzhaf I, Lenz T. Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist
blood pressure monitor for use on adults versus auscultation as defined by ANSI/AAMI SP 10-1992 Blood Pres Monit 2000;5:239-245 |
| 2 |
Nolly H, Osso P, Nolly M. Home blood pressure monitoring Intercontinental Cardiology 2001;10:111-114 |
| 3 |
Nolly H, Osso A, Romero M, Nolly M. The reproducibility of home- self measurements using the Braun PrecisionSensor BP 2550 Monitor J Hypertens 2001;19 (suppl 2):S26 |
| 4 |
Nolly H, Romero M, Nolly A, Nolly M. Home blood pressure (BP) measurement using B.P. Braun Vitalscan monitor. J Hypertens 200;18(suppl 4):S119 (Abstract) |
| 5 |
Sandevik E, Dyrdal A. Evaluation of the Braun BP2550 PrecisionSensor wrist blood pressure monitor in clinical practice [dabl® Educational Trust regrets that full details are missing] (Abstract) |
| 6 |
Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002 Blood Pressure Monitoring 2006;11:91-95 |
| 7 |
Nolly H, Romero M, Nolly A, Osso P, Reinoso O, Nolly M. Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol Blood Press Monit 2004;9:53-58 |
| 8 |
Nolly H, Osso P, Nolly A, Silva M, Nolly M, Romero M. Self-measurement of blood pressure: validation of the Braun BP VitalScan Plus 1650 monitor according to the European Society of Hypertension International Protocol Blood Press Monit 2008;13:101-106 |
| 9 |
Cotte UV, Faltenbacher VH, von Willich W, Bogner JR. Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656C Blood Press Monit 2008;13:55-62 |
| 10 |
Altunkan Ş, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in the elderly Blood Press Monit 2006;11:97-102 |
| 11 |
Yarrows SA. Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled American Journal of Hypertension 2004;17(No 1):54-58 |
| 12 |
Altunkan Ş, Öztas K, Altunkan E. Validation of the Omron 637IT wrist blood pressure measuring device with a position sensor according to the International Protocol in adults and obese adults Blood Press Monit 2006;11:79-85 |
| 13 |
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT) Blood Press Monit 2006;11:165-171 |
| 14 |
Angeli F, Sardone M, Angeli E, Repaci S, Gattobigio R, Verdecchia P. Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure. Blood Press Monit 2006; 11:349-354 |
| 15 |
Nolly H, Romero M, Nolly A, Silva M, Nolly M. Clinical evaluation of the Braun BP 3000 Easy Click according to the ANSI/AAMI SP 10:2002. Blood Press Monit 2006;11:91-95 |
| 16 |
Iliman N, Altunkan S, Kayaturk N, Altunkan E. Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an EasyClick cuff according to the International Protocol in adults. Blood Press Monit 2007;12:45-49 |
| 17 |
Altunkan S, Iliman N, Altunkan E. Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International Protocol Blood Press Monit 2007;12:119-125 |
| 18 |
Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G. Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension Blood Press Monit 2007;12:233-242 |
| 19 |
Zaetta V, Longo D, Perkovic D, Perfetti P, Gabrieli A, Pratticò F, Winnicki M. Validation of the SAW-102 wrist home blood pressure monitor according to the protocols of the British Hypertension Society, the Association for the Advancement of Medical Instrumentation, and the European Society of Hypertension Blood Press Monit 2009;14:32-36 |
| 20 |
El-Assaad MA, Orobinskaya LV, Topouchian JA, Asmar RG. Validation of the Omron® RX-3 Blood Pressure Measuring Device at the wrist level according to the International Protocol of the European Society of Hypertension AJH 2005;18(No. 5, Part 2):29A (P58) |
| 21 |
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of the Microlife BP W200-1 wrist device for blood pressure measurement. Blood Press Monit 2008;13:295-298 |
| 22 |
Dorigatti F, Bonso E, Saladini F, Palatini P. Validation of the visocor HM40 wrist blood pressure measuring device according to the International Protocol Blood Press Monit 2009;14:83-86 |
| 23 |
Palatini P, Dorigatti F, Bonso E, Ragazzo F. Validation of Microlife BP W100 wrist device assessed according to the European Society of Hypertension and the British Hypertension Society protocols Blood Press Monit 2009;14:41-44 |
| 24 |
Sheng CS, Huang QF, Wang J, Zhang Y, Li Y, Wang JG. Validation of the Health & Life HL168JD wrist blood pressure monitor for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2009;14:125-127 |
| 25 |
Germano G, Psimenos A, Sarullo F, Venditti A, Pecchioli V, Asmar R. Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol: The Pic Indolor Personal Check, Comfort Check, My Check and Travel Check. Blood Pressure 2009;18 (Suppl 1):15-23 |
| 26 |
Asmar R, Khabouth J, Topouchian J, El Feghali R, Mattar J. Validation of three automatic devices for self-measurement of blood pressure according to the International Protocol: The Omron M3 Intellisense (HEM-7051-E), the Omron M2 Compact (HEM 7102-E), and the Omron R3-I Plus (HEM 6022-E). Blood Press Monit 2010;15(1):49-54 doi: 10.1097/MBP.0b013e3283354b11 |
| 27 |
Saladini F, Benetti E, Palatini P. Accuracy of the visomat handy wrist blood pressure measuring device according to the International Protocol. Blood Press Monit 2010;15(5):281-284 doi: 10.1097/MBP.0b013e32833e50f2 |
| 28 |
Tian HY, Liu WJ, Li SG, Song Z, Gong W. Validation of the Transtek TMB-988 wrist blood pressure monitor for home blood pressure monitoring according to the International Protocol. Blood Press Monit 2010;15(6):326-328 doi: 10.1097/MBP.0b013e32833f56fb |
| 29 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol. Blood Press Monit. Blood Press Monit 2012;17(5):42-4 doi: 10.1097/MBP.0b013e32834f8301 |
| 30 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press Monit 2012;17(5):204-6 doi: 10.1097/MBP.0b013e328359c426 |
| 31 |
de Greeff A, Shennan AH. Validation of the Nissei WSK-1011 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2011 Sep 09 9 p. Available from: http://www.dableducational.org/Publications/2011/ESH-IP 2010 Validation of Nissei WSK-1011.pdf. |
| 32 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Kingyield BP210 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Jan 07. 7 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Kingyield BP210.pdf |
| 33 |
Zeng WF, Huang QF, Sheng CS, Li Y, Wang JG. Validation of the Rossmax S150 wrist blood pressure monitor, for self measurement, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Apr 08. 8 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Rossmax S150.pdf |
| 34 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON RS3 (HEM-6130-E) wrist blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01. 4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron RS3 (HEM-6130-E).pdf |
| 35 |
Takahashi H, Yokoi T, Yoshika M. Validation of the OMRON RS6 (HEM-6221-E) wrist blood pressure monitor, in oscillometry mode, for clinic use and self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Feb 01.4 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of Omron RS6 (HEM-6221-E).pdf |
| 36 |
Shag F, Zhu Z, Liu L, Wan Y. Validation of the iHealth BP7 (KD-972) wrist blood pressure monitor, in oscillometry mode, for self measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010 [Internet]. Dublin: dablEducational Trust; 2013 Mar 25. 9 p. Available from: http://www.dableducational.org/Publications/2013/ESH-IP 2010 Validation of iHealth BP7 (KD-972).pdf |
Devices for Measuring Blood Pressure in the Community
For a complete list of devices see our Device Index and for discontinued devices see our Discontinued Devices Index.
New to the website |
|
|
| Device |
Mode |
AAMI |
BHS |
ESH 2002 |
ESH 2010 |
Circumstance |
Recommendation Ref |
|
| A&D TM-2655 |
Osc |
Pass |
A/A |
|
|
Methods poorly explained |
Questionable 3 |
|
| Keito |
Wrist |
|
|
|
|
Standing; Fail Ad Hoc protocol |
Questionable 2 |
|
| Pharma-Smart PS-2000 |
Cuff |
Pass |
A/A |
|
|
At rest; BHS protocol violations |
Questionable 1 |
|
References for Validation of Devices for Measuring Blood Pressure in the Community
| 1 |
Alpert BS. Validation of the Pharma-Smart PS-2000 public use blood pressure monitor Blood Press Monit 2004;9:19-23 |
| 2 |
Staal EM, Nygard OK, Omvik P, Gerdts E. Blood pressure measurements by the Keito machine. Evaluation versus office blood pressure by physicians Blood Press Monit 2004;9:167-172 |
| 3 |
Kobalava ZD, Kotovskaya YV, Babaeva LA, Moiseev VS. Validation of oscillometric device for blood pressure measurement Blood Press Monit 2006;11:87-90 |
|